Thickness effects on maximum von-Mises stress of a cement mantle in total hip replacement - a finite element study
Y S Arun Kumar1, Bhaskar Pant, Konjengbam Darunkumar Singh
1Department of Civil Engineering, Indian Institute of Technology, Guwahati - India.
Aim:
The present study seeks to increase the life term of fully cemented total hip replacements by minimizing the stress values within the cement mantle.
Methods:
Three-dimensional (3D) finite element analyses have been carried out to investigate the effects of varying cement thickness on the von-Mises stress of a cement mantle. The magnitude and location of maximum von-Mises stress within the cement mantle have been studied for both straight and tapered prosthetic stems.
Results:
For prosthetic stems having lower radii sizes, the maximum stress zone is found in the upper region of the cement mantle whilst for stems with higher radii sizes the maximum stress zone is found in the lower region of the cement mantle. For the same cement thickness, straight stems are found to produce lower maximum stress values in the cement when compared to tapered stems. Finally, for the straight models with the same cement thickness, maximum stress values are found to decrease with increasing stem radius.
Conclusions:
It can be concluded that the maximum stress values in the cement mantle decrease with decreasing cement thickness.
More Related Videos
Related Concept Videos
Tensile Strength Considerations of Concrete
The dimensions and shape of a concrete specimen also...
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each material's...
Stress Concentrations in Circular Shafts
Temperature Dependent Deformation
Dynamic Modulus of Elasticity of Concrete
The sonic test is a common method to determine the dynamic modulus. In this test, a concrete beam, sized either 6 x 6 x 30 inches or 4 x 4 x 20 inches, is clamped at its center. Vibrations are initiated at one end of the beam by an electromagnetic exciter unit powered by a...