Does change in thoracic impedance measured via defibrillator electrode pads accurately detect ventilation breaths in
Kathryn Roberts1, Vijay Srinivasan, Dana E Niles
1Department of Nursing, The Children's Hospital of Philadelphia, Philadelphia, PA 19104, USA.
Insights
Changes in thoracic impedance (ΔTI) can accurately detect adequate rescue ventilations (≥ 7 ml/kg) in children. Both anterior-apical and anterior-posterior defibrillator pad positions showed similar detection rates for effective ventilation volumes.
Area of Science:
- Pediatric critical care medicine
- Biomedical engineering
- Cardiopulmonary resuscitation
Background:
- Current resuscitation guidelines recommend rescue ventilation tidal volumes of 7-10 ml/kg.
- Changes in thoracic impedance (ΔTI) have not been previously studied in children for detecting ventilation adequacy.
Purpose of the Study:
- To evaluate if ΔTI measured using standard anterior-apical (AA) defibrillator pad positions can accurately detect ventilations > 7 ml/kg in children.
- To compare the efficacy of the standard AA position with the alternative anterior-posterior (AP) position for ΔTI measurements.
Main Methods:
- Prospective, observational study involving 28 sedated children (6 months to 17 years) on mechanical ventilation.
- Thoracic impedance (TI) was measured using a Philips MRx defibrillator with standard electrode pads in both AA and AP positions.
- Ventilation volumes were simultaneously measured using a pneumotachometer for comparison.
Main Results:
- For ventilations ≥ 7 ml/kg, detection rates were high and similar for both positions: 95.1% for AA and 95.7% for AP (p=0.38).
- For ventilations < 7 ml/kg, detection rates were lower (80.0% for both positions), indicating potential missed shallow breaths.
- No significant difference in impedance measurements was found between the AA and AP pad positions.
Conclusions:
- Thoracic impedance changes measured via defibrillator pads can reliably detect adequate (≥ 7 ml/kg) ventilation volumes in mechanically ventilated children.
- Both AA and AP pad positions demonstrate similar accuracy, but may miss smaller ventilation volumes (< 7 ml/kg).
- The findings support the use of ΔTI as a tool to guide rescue ventilations during cardiopulmonary resuscitation (CPR) in pediatric patients.
Introduction:
Resuscitation guidelines recommend rescue ventilations consist of tidal volumes 7-10 ml/kg. Changes in thoracic impedance (ΔTI) measured using defibrillator electrode pads to detect and guide rescue ventilations have not been studied in children.
Aim:
We hypothesized that ΔTI measured via standard anterior-apical (AA) position can accurately detect ventilations with volume > 7 ml/kg in children. We also compared standard AA position with alternative anterior-posterior (AP) position.
Methods:
IRB-approved, prospective, observational study of sedated, subjects (6 months to 17 years) on conventional mechanical ventilation. Thoracic impedance (TI) was obtained via Philips MRx defibrillator with standard electrode pads for 5 min each in AA and AP positions. Ventilations were simultaneously measured by pneumotachometer (Novametrix CO(2)SMO Plus).
Results:
Twenty-eight subjects (median 4 years, IQR 1.7-9 years; median 16.3 kg, IQR 10.5-39 kg) were enrolled. Data were available for 21 episodes in AA position and 22 episodes in AP position, with paired AA and AP data available for 18. For ventilations with volume < 7 ml/kg, the defibrillator algorithm detected 80.0% for both AA and AP (p=0.99). For ventilations ≥ 7 ml/kg, detection was 95.1% for AA and 95.7% for AP (p=0.38).
Conclusions:
Changes in thoracic impedance obtained via defibrillator pads can accurately detect ventilations above 7 ml/kg in stable, mechanically ventilated children, corresponding to rescue ventilations recommended during CPR. Both AA and AP pad positions were less sensitive to detect smaller volumes (< 7 ml/kg) than higher volumes (≥ 7 ml/kg), suggesting that shallow ventilations during CPR might be missed. There were no differences in impedance measurements between standard AA pad position and commonly used alternative AP pad position.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Cardiopulmonary Resuscitation III: AED Use
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Physical Assessment of the Respiratory Tract II: Palpation
Thoracic Palpation
Thoracic palpation detects tenderness, masses, lesions, respiratory excursions, and vocal fremitus. The nurse assesses...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.