Quantification of lumen stenoses with known dimensions by conventional angiography and computed tomography:
A Arbab-Zadeh1, J Texter, K M Ostbye
1Department of Medicine, Division of Cardiology, Johns Hopkins University, Baltimore, Maryland, USA. ArminZadeh@jhu.edu
Background:
Quantitative coronary angiography (QCA) has inherent limitations for displaying complex vascular anatomy, yet it remains the gold standard for stenosis quantification.
Objective:
To investigate the accuracy of stenosis assessment by multi-detector computed tomography (MDCT) and QCA compared to known dimensions.
Methods:
Nineteen acrylic coronary vessel phantoms with precisely drilled stenoses of mild (25%), moderate (50%) and severe (75%) grade were studied with 64-slice MDCT and digital flat panel angiography. Fifty-seven stenoses of circular and non-circular shape were imaged with simulated cardiac motion (60 bpm). Image acquisition was optimised for both imaging modalities, and stenoses were quantified by blinded expert readers using electronic callipers (for MDCT) or lumen contour detection software (for QCA).
Results:
Average difference between true and measured per cent diameter stenosis for QCA was similar compared to MDCT: 7 (+/-6)% vs 7 (+/-5)% (p=0.78). While QCA performed better than MDCT in stenoses with circular lumen (mean error 4 (+/-3)% vs 7 (+/-6)%, p<0.01), MDCT was superior to QCA for evaluating stenoses with non-circular geometry (mean error 10 (+/-7)% vs 7 (+/-5)%, p<0.05). In such lesions, QCA underestimated the true diameter stenosis by >20% in 9 of 27 (33%) vs 1 of 29 (3%) in lumen with circular geometry.
Conclusions:
QCA often underestimates diameter stenoses in lumen with non-circular geometry. Compared to QCA, MDCT yields mildly greater measurement errors in perfectly circular lumen but performs better in non-circular lesions. These findings have implications for using QCA as the gold standard for stenosis quantification by MDCT.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT