Oxygen delivery during transtracheal oxygenation: a comparison of two manual devices
François Lenfant1, Didier Péan, Laurent Brisard
1Department of Anesthesiology, CHU de Dijon, General Hospital, Dijon Cedex, France. francois.lenfant@chu-dijon.fr
Background:
The Manujet™ and the ENK Oxygen Flow Modulator™ (ENK) deliver oxygen during transtracheal oxygenation. We sought to describe the ventilation characteristics of these 2 devices.
Methods:
The study was conducted in an artificial lung model consisting of a 15-cm ringed tube, simulating the trachea, connected via a flow analyzer and an artificial lung. A 15-gauge transtracheal wire reinforced catheter was used for transtracheal oxygenation. The ENK and Manujet were studied for 3 minutes at respiratory rates of 0, 4, and 12 breaths/min, with and without the artificial lung, in a totally and a partially occluded airway. Statistical analysis was performed using analysis of variance followed by a Fisher exact test; P < 0.05 was considered significant.
Results:
Gas flow and tidal volume were 3 times greater with the Manujet than the ENK (approximately 37 vs 14 L · min(-1) and 700 vs 250 mL, respectively) and were not dependent on the respiratory rate. In the absence of ventilation, the ENK delivered a 0.6 ± 0.1 L · min(-1) constant gas flow. In the totally occluded airway, lung pressures increased to 136 cm H(2)O after 3 insufflations with the Manujet, whereas the ENK, which has a pressure release vent, generated acceptable pressures at a low respiratory rate (4 breaths/min) (peak pressure at 27.7 ± 0.7 and end-expiratory pressure at 18.8 ± 3.8 cm H(2)O). When used at a respiratory rate of 12 breaths/min, the ENK generated higher pressures (peak pressure at 95.9 ± 21.2 and end-expiratory pressure at 51.4 ± 21.4 cm H(2)O). In the partially occluded airway, lung pressures were significantly greater with the Manujet compared with the ENK, and pressures increased with the respiratory rate with both devices. Finally, the gas flow and tidal volume generated by the Manujet varied proportionally with the driving pressure.
Discussion:
This study confirms the absolute necessity of allowing gas exhalation between 2 insufflations and maintaining low respiratory rates during transtracheal oxygenation. In the case of total airway obstruction, the ENK may be less deleterious because it has a pressure release vent. Using a Manujet at lower driving pressures may decrease the risk of barotrauma and allow the safe use of higher respiratory rates.
More Related Videos
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Tracheostomy Suctioning II: Procedure