Related Experiment Video
Updated: Jun 9, 2026

Infrared Degenerate Four-wave Mixing with Upconversion Detection for Quantitative Gas Sensing
Published on: March 22, 2019
Tunable 17-microm laser spectrometer for optical absorption measurements of CH(4), C(2)H(4), and high-temperature HCl
Abstract:
A tunable 1.7-microm cw F(2)(+)-center laser spectrometer system that is continuously tunable from 1.43 to 1.78 microm in wavelength and has aspectral linewidth (FVWM) of 0.07 cm(-1) is developed. This system is used to measure the 2-0 rotational-vibrational absorption line profiles of hydrogen chloride (HCl) at high temperatures, the Boltzmann thermal equilibrium temperature, and to determine the extent of potential interference or overlap of the HCl lines with those that are due to hot water vapor. As a demonstration of the utility of the laser spectrometer system, it is also used to measure the spectra of methane (CH(4)) and ethylene (C(2)H(4)) near 1.65 microm, and is routed through a fiber-optic cable to a remote site to detect a CH(4) plume.
Related Concept Videos
UV–Vis Spectrometers
IR Spectrometers
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...