Related Experiment Video
Updated: Jun 9, 2026

Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
[Dose reduction at CT of the lumbar spine using a 320-detector row scanner: initial results]
1Service d'Imagerie Guilloz, Hôpital Central, CHU Nancy, 29 avenue du Maréchal de Lattre de Tassigny, 54035 Nancy cedex. alban.gervaise@hotmail.fr
Purpose:
To compare radiation dose and image quality for CT of the lumbar spine between helical CT and wide volume mode scanning with a 320-detector row CT.
Patients And Methods:
Monocenter prospective study on 20 consecutive patients divided into two groups. All 20 patients underwent lumbar spine CT on the 320-detector row scanner (Aquilion One, Toshiba). The CT examinations for group 1 were performed using the wide volume mode with 320 detector rows while the CT examinations for group 2 were performed using a 64-detector row helical CT mode. The acquisition length and delivered dose corresponding to the DLPe (extended dose length product) as well as qualitative and quantitative image quality were compared between both groups.
Results:
The mean acquisition length was comparable between both groups. There was a significant dose reduction of about 35% for group 1 compared to group 2 (mean DLPe of 970 mGy.cm for group 1 compared to 1503 mGy.cm for group 2, p<0.028) when using the wide volume mode acquisition at 320-detector row CT compared to the 64-detector row helical CT mode. No significant difference was noted for image quality between both groups.
Conclusion:
The acquisition of lumbar CT using the wide volume mode at 320-detector row CT allows significant dose reduction to patients compared to the 64-detector row helical CT mode while preserving image quality.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
![Quantitative [18F]-Naf-PET-MRI Analysis for the Evaluation of Dynamic Bone Turnover in a Patient with Facetogenic Low Back Pain](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F58491.jpg&w=3840&q=50)