Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Dose verifications by use of liquid ionization chamber of an electronic portal imaging device (EPID)
Kunihiko Tateoka1, Atsushi Oouchi, Kensei Nakata
1Radiation Oncology, Imaging and Diagnosis, Division of Organ Function and Therapeutics, Molecular and Organ Regulation, Sapporo Medical University Graduate School of Medicine, Sapporo, Japan. tateoka@sapmed.ac.jp
Abstract:
In this study, we examined the ability of an L-EPID to verify rectangular and irregular fields and to measure the transmitted exit doses. With respect to the beam profile of rectangular and irregular fields and the doses transmitted through an inhomogeneous phantom, the L-EPID dose obtained from the L-EPID measurement was compared with the conventional dose measured by use of a 0.12-cc ionization chamber and a 3D water phantom. In the comparison of the rectangular and irregular fields, the difference in the off-center ratio (OCR) between the L-EPID dose and the conventional dose was approximately 3% in the steep-dose-gradient region (penumbra regions, >30%/cm) and approximately +/-0.5% in the gentle-dose-gradient region (5%/cm). On the other hand, the dose differences between the L-EPID and the measured doses were less than approximately 2% in the gentle-dose-gradient region. In addition, in the steep-dose-gradient region, the maximum difference was 30%. However, the differences in the distance-to-agreement (DTA) were less than approximately +/-1 mm and were unrelated to the dose gradient. These results suggest that dose verification by L-EPID is very useful in clinical applications.
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Gas Chromatography: Types of Detectors-II
Ion-Exchange Chromatography