Related Experiment Video
Updated: Jun 9, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Evaluation of mean glandular dose in a full-field digital mammography unit in Tabriz, Iran
Hamed Alizadeh Riabi1, Parinaz Mehnati, Asghar Mesbahi
1Department of Medical Physics, Faculty of Medicine, Tabriz University of Medical Sciences, Tabriz, Iran.
Abstract:
This study was aimed at evaluating the mean glandular dose (MGD) and affecting factors during mammography examinations by a full-field digital mammography unit. An extensive quality control program was performed to assure that the unit is properly working. Required information including compressed breast thickness (CBT), breast parenchymal pattern and technical factors used for imaging were recorded. An entrance skin exposure measurement was also performed using slabs of polymethylmethacrylate with 2-8 cm thickness. On the basis of recorded information and measured data, the MGD was estimated for 1145 mammography examinations obtained from 298 patients. Mean CBTs of 4.9 and 5.8 cm and MGDs of 2 and 2.4 mGy were observed for craniocaudal and mediolateral oblique views, respectively. Significant correlation was seen between MGD and CBT, breast parenchymal pattern and applied kVp and mAs.

