Comparison of electronic portal imaging and cone beam computed tomography for position verification in children
M S Zaghloul1, A G Mousa, E Eldebawy
1Radiation Oncology Department, Children's Cancer Hospital, Cairo, Egypt. mszagh@yahoo.com
Aim:
To compare the accuracy of radiotherapy set-up using an electronic portal imaging device (EPID) versus megavoltage cone beam computed tomography (MV-CBCT) in paediatric patients.
Materials And Methods:
In total, 204 pairs of EPID and MV-CBCT were carried out for 72 patients in the first 3 treatment days and weekly thereafter.
Results:
For the whole group, the mean systematic EPID set-up errors were 1.8 (±1.7), 1.6 (±1.3), 1.4 (±1.5) mm and 2.3 (±1.7), 1.6 (±1.3), 2.4 (±1.6) mm for MV-CBCT in the longitudinal, lateral and vertical directions, respectively, whereas the mean EPID random errors were 2.0 (±1.7), 1.4 (±1.5), 1.2 (±1.6) and 1.9 (±1.5), 1.5 (±1.3), 2.1 (±1.7) mm for MV-CBCT in the longitudinal, lateral and vertical directions, respectively. For systematic errors of head and neck patients, there was a statistically significant difference in the lateral and vertical directions (P=0.027, 0.003), whereas in the non-head and neck patients there was a statistically significant difference in the lateral direction only (P=0.031). In head and neck patients, the mean random errors were significantly different in the vertical and lateral directions, whereas in non-head and neck patients, they were significantly different in the vertical direction only. The larger values alternate between the two modalities. The systematic and random errors (detected by EPID and MV-CBCT) were significantly correlated in almost all direction in all tumour sites.
Conclusions:
The comparison between set-up error in EPID and MV-CBCT was not in favour of any of the two modalities. However, the two modalities were strongly correlated but fairly agreed and the differences between the shifts reported were small and hardly influenced the recommended planning target volume margin.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...