Related Experiment Video
Updated: Jun 8, 2026

A Novel Technique for Raman Analysis of Highly Radioactive Samples Using Any Standard Micro-Raman Spectrometer
Published on: April 12, 2017
In-field radon measurement in water: a novel approach
S A Talha1, R J de Meijer, R Lindsay
1Department of Physics, University of the Western Cape, Bellville 7535, South Africa. siddig.talha@gmail.com
Abstract:
This paper presents a novel approach of measuring radon in-water in the field by inserting a MEDUSA gamma-ray detector into a 210 L or 1000 L container. The experimental measurements include investigating the effect of ambient background gamma-rays on in-field radon measurement, calibrating the detector efficiency using several amounts of KCl salt dissolved in tap water, and measuring radon in borehole water. The results showed that there is fairly good agreement between the field and laboratory measurements of radon in water, based on measurements with Marinelli beakers on a HPGe detector. The MDA of the method is 0.5 Bq L⁻¹ radon in-water.