Comparative study of two generations of the Complior device for aortic pulse wave velocity measurements
Telmo Pereira1, João Maldonado
1Escola Superior de Tecnologia da Saúde de Coimbra, Departamento de Cardiopneumologia, Instituto de Investigação e Formação Cardiovascular, Coimbra, Portugal. telmo@estescoimbra.pt
Introduction:
The purpose of this study was to evaluate intraobserver and interobserver reproducibility of carotid-femoral pulse wave velocity (PWV) with the Complior method, by comparing two generations of the device and the manual method.
Methods:
We studied 31 participants (9 women) with a mean age of 47.67±12.55 years, who were submitted to sequential measures of carotid-femoral PWV by two experienced operators. The evaluations were made by randomly switching between the manual method, the Complior first generation, and the Complior SP (third generation).
Results:
The analysis of concordance of measurements of PWV showed strong intraobserver and interobserver correlations. The intraobserver correlation coefficients were 0.984, 0.977, and 0.966, respectively, for manual evaluation, the Complior first generation, and the Complior SP (P<0.0001). In the case of interobserver reproducibility, we found correlations of 0.940, 0.979, and 0.974, respectively, for manual evaluation, the Complior first generation, and Complior SP (P<0.0001). Good indicators of reproducibility were also a withdrawal of the Bland-Altman analysis, with most values within 2 standard deviations. The mean difference±SD for intraobserver and interobserver agreement was -0.006±0.05 and -0.06±0.09 for manual determination; -0.02±0.06 and -0.04±0.06 for Complior first generation; and 0.16±0.09 and 0.15±0.07 for the Complior SP.
Conclusion:
The data showed a high reproducibility of the Complior method, documented by excellent correlations and low mean differences in measurements within and between observers. These data showed unequivocally that this method met the quality requirements for its inclusion in integrated clinical follow-up programs.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation