Evaluation of tracheal intubation in several positions by the Pentax-AWS Airway Scope: a manikin study
Nobuyasu Komasawa1, Ryusuke Ueki, Motoi Itani
1Department of Anesthesiology, Hyogo College of Medicine, Mukogawa-cho 1-1, Nishinomiya, Hyogo 663-8501, Japan. nkomazaw@hyo-med.ac.jp
Purpose:
In emergency situations, rescuers must occasionally secure the airway while the patient is in a restricted position rather than in the ideal supine position. We hypothesized that the Pentax-AWS Airway Scope (AWS) may be useful for emergent tracheal intubation in such positions.
Methods:
Thirteen non-anesthesia residents performed tracheal intubation on a simulated manikin in the supine (Supine), left-lateral decubitus (Left-LT), right-lateral decubitus (Right-LT), prone (Prone), and sitting (Sitting) position, respectively, to assess AWS performance.
Results:
Intubations were successful in all five positions. The time needed to secure the airway did not differ significantly between the Left-LT and Supine positions. Doctors required significantly more time to secure the airway in the Prone, Sitting, and Right-LT positions than in the Supine position. Visual analog scale (VAS) scoring of the subjective difficulty of laryngoscopy was lower in the Supine position rather than in the Right-LT, Prone, and Sitting positions. The VAS score of subjective difficulty of tracheal tube passage through the glottis was significantly higher in the Sitting position than in the other four positions.
Conclusion:
Although tracheal intubations with AWS in all five positions tested were successful, intubation with the patient in the Sitting, Right-LT, and Prone positions was more difficult and required more time than that in the Supine position.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...