Background:
Highly cross-linked polyethylene (HXLPE), created by disrupting the molecular structure of polyethylene, then through the application of heat, encourages creation of new cross-links in the process, resulting in a material with improved wear resistance. The impetuses for this new technology were the unsatisfactory wear properties and subsequent osteolysis of noncross-linked polyethylene. A 72% reduction in wear using highly cross-linked polyethylenes (HXLPE) compared with conventional polyethylene at 5 years was described previously. The longest term followup studies on HXLPE range from 2 to 6 years.
Questions/Purposes:
We therefore addressed the following questions: (1) Does the improvement in wear observed at the earlier followup continue to 7 to 10 years? (2) What is the incidence of osteolysis in this group of patients and in the control group?
Methods:
We retrospectively reviewed 38 prospectively followed patients who had 42 hips with an annealed HXLPE who were followed a minimum of 7 years (average, 8.6 years; SD=1; range, 7-10.3 years). Wear and osteolysis were compared with those of a control group of 39 patients (40 hips) from a US Investigational Device Exemption (IDE) prospective, randomized study begun in 1996 with conventional polyethylene and followed for a minimum of 6 years (average, 7.5 years; SD=1.1; range, 6-10.2 years). Linear head penetration was measured from AP radiographs at early, 1-year, 5-year, and most recent followups.
Results:
At the average followup, annual linear wear was 0.031 mm (SD=0.014) for the HXLPE and 0.141 mm (SD=0.080) for the control group, a 78% reduction. No mechanical failure of the polyethylene was noted in either group. Incidence of osteolysis was 50% in the control group (all lesions confined to proximal Gruen Zones 1 and 7) compared with no cases in the investigational group.
Conclusions:
We observed an improvement in wear and no mechanical failures with this annealed material.
Level Of Evidence:
Level III, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
Related Concept Videos
Polymer Classification: Architecture
Plastic Behavior
Free-Radical Chain Reaction and Polymerization of Alkenes
Abrasion Resistance of Concrete
One such test is the revolving disc test, where three plates...
Olefin Metathesis Polymerization: Acyclic Diene Metathesis (ADMET)
Similar to cross-metathesis, ADMET also involves the formation of metallacyclobutane intermediate by [2+2] cycloaddition of one of the double bonds of a terminal diene with...
Ziegler–Natta Chain-Growth Polymerization: Overview