User experience network. Supply gas failure alarm on Cardinal Health Infant Flow SiPAP units may not activate
Abstract:
The supply gas failure alarm on Cardinal Health Infant Flow SiPAP units manufactured before April 2009 may not activate in the event of a gas supply loss if the device's silencer accessory is attached. However, the unit's FiO2 (fraction of inspired oxygen) and low-airway-pressure alarms will activate in such cases. If both of these alarms activate simultaneously, users should suspect a failure of the gas supply pressure. Identifying affected units requires testing that can be conducted during the device's next scheduled maintenance.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Errors occurring during blood pressure monitoring
Several factors...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...