Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length
Paul D Deak1, Yulia Smal, Willi A Kalender
1Institute of Medical Physics, University Erlangen-Nürnberg, Henkestr. 91, Erlangen, Germany.
Purpose:
To determine conversion factors for the new International Commission on Radiological Protection (ICRP) publication 103 recommendations for adult and pediatric patients and to compare the effective doses derived from Monte Carlo calculations with those derived from dose-length product (DLP) for different body regions and computed tomographic (CT) scanning protocols.
Materials And Methods:
Effective dose values for the Oak Ridge National Laboratory phantom series, including phantoms for newborns; 1-, 5-, and 10-year-old children; and adults were determined by using Monte Carlo methods for a 64-section multidetector CT scanner. For each phantom, five anatomic regions (head, neck, chest, abdomen, and pelvis) were considered. Monte Carlo simulations were performed for spiral scanning protocols with different voltages. Effective dose was computed by using ICRP publication 60 and publication 103 recommendations. The calculated effective doses were compared with those derived from the DLP by using previously published conversion factors.
Results:
In general, conversion factors determined on the basis of Monte Carlo calculations led to lower values for adults with both ICRP publications. Values up to 33% and 32% lower than previously published data were found for ICRP publication 60 and ICRP publication 103, respectively. For pediatric individuals, effective doses based on the Monte Carlo calculations were higher than those obtained from DLP and previously published conversion factors (eg, for chest CT scanning in 5-year-old children, an increase of about 76% would be expected). For children, a variation in conversion factors of up to 15% was observed when the tube voltage was varied. For adult individuals, no dependence on voltage was observed.
Conclusion:
Conversion factors from DLP to effective dose should be specified separately for both sexes and should reflect the new ICRP recommendations. For pediatric patients, new conversion factors specific for the spectrum used should be established.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Biological Effects of Radiation
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Dose Response Curve: Conventional Versus Nonmonotonic
Imaging Studies for Cardiovascular System V: CT