Influence of ocular pulse amplitude on ocular response analyzer measurements
Guihua Xu1, Dennis Shun Lam, Christopher Kai-shun Leung
1Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Hong Kong, The People's Republic of China.
Purpose:
To evaluate if the ocular pulse amplitude (OPA) (the difference between systolic and diastolic intraocular pressure) is associated with the measurement variability of IOPg [Goldmann-correlated intraocular pressure (IOP)], IOPcc (corneal compensated IOP), corneal hysteresis (CH), and corneal resistance factor (CRF) obtained from the ocular response analyzer (ORA).
Methods:
Sixty normal participants and 60 glaucoma patients were included. In 1 randomly selected eye in each participant, 4 repeated measurements were obtained with the ORA (Reichert Inc, Depew, NY) followed by OPA measurement by dynamic contour tonometry (Pascal; Swiss Microtechnology AG, Port, Switzerland). The repeatability of IOPg (a mathematically derived value with strong correlation with Goldmann applanation pressure measurement), IOPcc (a mathematically derived measurement less affected by corneal biomechanical properties), CH, and CRF were calculated. The associations between age, refraction, keratometry, central corneal thickness, axial length, OPA, and ORA measurements variability were evaluated with univariate and multivariate regression analyses.
Results:
The repeatability of IOPg, IOPcc, CRF, and CH were 4.08 (95% confidence interval: 3.06-5.09) mm Hg, 4.72 (3.54-5.89) mm Hg, 2.14 (1.61-2.68) mm Hg, 2.35 (1.77-2.94) mm Hg, respectively, for the normal group; and 4.18 (3.75-4.61) mm Hg, 5.56 (4.99-6.14) mm Hg, 2.17 (1.95-2.40) mm Hg, 2.78 (2.50-3.07) mm Hg, respectively, for the glaucoma group. Within-subject variances of IOPg and IOPcc, but not CRF and CH, were positively correlated with OPA. No association was found between age, refraction, keratometry, central corneal thickness, axial length, and the measurement variability of ORA.
Conclusions:
The measurement reliability of ORA was only moderate. Eyes with large OPA were associated with high IOP measurement variability. Taking average of multiple repeated measurements is important for reliable measurement of ORA.
Related Concept Videos
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring pulse
Pulse
The pulse serves as a clinical indicator...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...