Neuroprotection during carotid artery stenting using the GORE flow reversal system: 30-day outcomes in the EMPiRE

Daniel G Clair1, L Nelson Hopkins, Manish Mehta

  • 1Department of Vascular Surgery, Heart and Vascular Institute, Cleveland Clinic Lerner College of Medicine, Cleveland, Ohio 44195, USA. CLAIRD@ccf.org

Insights

The GORE Flow Reversal System offers effective neuroprotection during carotid artery stenting (CAS), showing low stroke and death rates. This innovative device is safe for a wide range of patients undergoing CAS procedures.

Area of Science:

  • Cardiovascular Medicine
  • Interventional Neurology
  • Medical Devices

Background:

  • Carotid artery stenting (CAS) involves risks, with existing embolic protection devices having limitations.
  • The GORE Flow Reversal System is a novel, proximally placed device designed for distal neuroprotection during CAS.
  • This system reverses internal carotid artery blood flow to divert emboli away from the brain.

Purpose of the Study:

  • To investigate the safety and efficacy of the GORE Flow Reversal System in a prospective, multicenter, single-arm clinical study (EMPiRE Clinical Study).
  • To evaluate 30-day major adverse events (MAE) in high-surgical-risk patients undergoing CAS with the flow reversal system.
  • To compare the observed MAE rate against an objective performance criterion (OPC) derived from previous CAS studies.

Main Methods:

  • The study enrolled 245 high-surgical-risk patients with carotid stenosis undergoing CAS using the flow reversal system.
  • Outcomes assessed included a composite of stroke, death, myocardial infarction, and transient ischemic attack within 30 days post-procedure.
  • Patient demographics included a mean age of 70 years, with 32% symptomatic and 16% aged 80 years or older.

Main Results:

  • The overall MAE rate was 4.5% (11 patients), significantly lower than the OPC (P=0.002).
  • The combined stroke and death rate was 2.9%, with no major ischemic strokes reported.
  • Subgroup analyses for symptomatic, asymptomatic, and octogenarian patients met American Heart Association guidelines, with death and stroke rates ranging from 2.6% to 3%.

Conclusions:

  • The GORE Flow Reversal System demonstrated a low stroke and death rate, among the lowest reported in CAS trials.
  • The system proved to be safe and effective for neuroprotection during CAS.
  • The study suggests the flow reversal system benefits a broad spectrum of patients undergoing carotid artery stenting.