Laser Stark spectrometer for the measurement of ammonia in flue gas
Abstract:
The physical background of a laser Stark spectrometer dedicated to the measurement of ammonia slip through DeNo(x) reactors in power stations is treated. The dependence of the ammonia measurement on temperature and pressure variations is derived and verified experimentally. Selection of spectral lines with both a good absorption coefficient and a high sensitivity to the Stark effect, within the range of the CO(2) laser, has been carried out. If a (12)CO(2) laser is used at a temperature of 573 K and at atmospheric pressure, the 10R8 laser line is recommended for best results. The 10R18 line of the (13)CO(2) laser yields a still higher sensitivity (detection limit 0.4 ppm) for a moderate electric field. Theoretical predictions for the sensitivity of ammonia detection are compared with experimental data. Results of measurements in an industrial environment are presented.
More Related Videos
07:14Automated, High-resolution Mobile Collection System for the Nitrogen Isotopic Analysis of NOx
Published on: December 20, 2016
10:29Calibrated Passive Sampling - Multi-plot Field Measurements of NH3 Emissions with a Combination of Dynamic Tube Method and Passive Samplers
Published on: March 21, 2016
Related Concept Videos
Atomic Fluorescence Spectroscopy
Flame Photometry: Overview
Atomic Emission Spectroscopy: Instrumentation
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Spectroscopy: Absorption, Emission, and Fluorescence
Flame Photometry: Lab