Long-path absorption measurement of CO(2) with a Raman-shifted tunable dye laser
Abstract:
A laser long-path absorption system was developed with a hydrogen Raman shifter pumped by a tunable dye laser. The absorption spectrum of CO(2) in the 2-µm region was measured in the open air with a retroreflector or a hard target. The concentration of CO(2) was determined from the spectrum by the least-squares method. Noise in the system was analyzed, and the propagation of error to the obtained concentration was investigated. The statistical error in the concentration was estimated at approximately 1% for a single spectrum measurement. Comparison with the simultaneous measurements with a nondispersive infrared gas analyzer showed good agreement.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
UV–Vis Spectroscopy: Beer–Lambert Law