Related Experiment Video
Updated: Jun 8, 2026

Employing the Forced Oscillation Technique for the Assessment of Respiratory Mechanics in Adults
Published on: February 9, 2022
Noninvasive high frequency oscillatory ventilation through nasal prongs: bench evaluation of efficacy and mechanics
Daniele De Luca1, Virgilio P Carnielli, Giorgio Conti
1Division of Neonatology and Institute of Mother and Child Health, G.Salesi Women and Children Hospital, Polytechnical University of Marche, Ancona, Italy. dm.deluca@fastwebnet.it
Background:
Noninvasive high frequency oscillatory ventilation through nasal prongs (nHFOV) has been proposed as a new respiratory support in neonatology. We studied the effect of ventilation parameters and nasal prongs on nHFOV efficacy and mechanics.
Methods:
Customized sealed circuits connecting a SM3100A oscillator to a neonatal lung model were developed to evaluate the effect of applying HFOV via two different sized nasal prongs on delivered tidal volume and pressure. Measurements were made across a range of frequencies and pressures; amplitude was set to obtain visible lung oscillation.
Results:
Volume delivered by peak-to-peak oscillation, ventilation, and pressure significantly differed among the interfaces, being higher for large cannulae and the control circuit (p < 0.0001). The interposition of a large or small nasal prong reduced volume to 56 and 26%, ventilation to 32 and 9%, and mean pressure to 83 and 79%, respectively, of the values measured for the direct connection of the oscillator to the test lung. Volume and ventilation were inversely related to frequency, which was particularly evident with larger diameter circuits due to higher delivered tidal volume (R (2) > 0.9). Increasing ventilation was associated with larger tidal volume and nasal prong diameter (adjusted R (2) = 0.97).
Conclusions:
nHFOV using common nasal prongs is technically possible. Efficiency of tidal volume delivery is significantly affected by prong diameter.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Suctioning the Nasopharyngeal Airway
Equipment Required
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: