Dual-source parallel RF transmission for clinical MR imaging of the spine at 3.0 T: intraindividual comparison with
Michael Nelles1, Roy S König, Jürgen Gieseke
1Department of Radiology, University of Bonn Hospital and Medical School, Sigmund Freud Strasse 25, D-53105 Bonn, Germany. michael.nelles@ukb.uni-bonn.de
Purpose:
To prospectively and intraindividually compare single-source radiofrequency (RF) excitation and dual-source parallel RF excitation in 3.0-T magnetic resonance (MR) imaging of the spine.
Materials And Methods:
Institutional review board approval and written informed patient consent were obtained. The RF power of a 3.0-T MR imaging system was distributed to two ports of the body coil of the system by using independent RF transmit channels. The maximum B(1) field strength for dual-source parallel RF excitation was maintained, as compared with single-source RF excitation. The repetition time was reduced according to the revised RF setup with dual-source parallel RF excitation while maintaining specific energy absorption limitations. Thirty patients were examined with and without dual-source parallel RF excitation. Diagnostic quality was assessed independently by two radiologists according to a four-point grading system. Image contrast ratios (CRs) were calculated between reference tissues and vertebrae for single-source RF excitation and dual-source parallel RF excitation.
Results:
The mean acceleration achieved with dual-source parallel RF excitation was 36% (range, 18%-50%). The total imaging duration of a three-station total spinal examination was reduced by one-third by using dual-source parallel RF transmission. For all cases investigated, diagnostic image quality without significant differences between the two methods and with a good interobserver agreement was achieved (Kendall tau-b, 0.50-0.84). The observed image contrast changes were predominantly small (<0.10 in 15 of 24 CRs), though they were significantly different (P < .05).
Conclusion:
While shortening examination times by approximately one-third, the dual-source parallel RF transmission mode in MR imaging of the spine yielded diagnostic image quality comparable to that with the conventional single-source RF transmission mode.