Related Experiment Video
Updated: Jun 8, 2026

Measurement of Ultrafast Vibrational Coherences in Polyatomic Radical Cations with Strong-Field Adiabatic Ionization
Published on: August 6, 2018
Miniaturized high-resolution time-of-flight mass spectrometer MULTUM-S II with an infinite flight path
Shuichi Shimma1, Hirofumi Nagao, Jun Aoki
1Center for Advanced Science and Innovation, Venture Business Laboratory, Osaka University, Suita, Osaka 565-0871, Japan. shuichi@mass.phys.sci.osaka-u.ac.jp
Abstract:
A new miniature multiturn time-of-flight (TOF) analyzer "MULTUM-S II" has been designed and constructed. This instrument consists of an electron ionization source, the multiturn TOF ion optics, a detector, vacuum system, and electronic circuits. The multiturn TOF analyzer consists of four electrostatic toroidal sectors and two additional electric toroidal sectors for the purpose of ion injection/ejection. The size and weight of the system is less than 50 cm × 57 cm × 30 cm and 35 kg (including vacuum pumps and electronic circuits). The multiturn TOF analyzer is capable of high mass resolution because of its infinite flight path utilizing perfect space and time focused closed flight orbit. To evaluate the resolution in MULTUM-S II, separation of pyridine (¹²C₅H₅N) and the isotopic component of benzene (¹³C¹²C₅H₆) was performed at a mass resolution of about 20,000. Another performance characteristic of the MULTUM-S II was demonstrated by the separation of the greenhouse gas doublet CO₂ and N₂O (Δm = 0.0113 Da). While the mass difference is a mere 0.01 Da, the instrument could easily separate the two peaks at a calculated mass resolution of 31,600. The MULTUM-S II offers high mass resolution mass spectrometry in a miniaturized/portable enclosure.
Related Concept Videos
Tandem Mass Spectrometry
Mass Analyzers: Common Types
MALDI-TOF Mass Spectrometry
Mass Spectrometers
Mass Analyzers: Overview
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...