Related Experiment Video
Updated: Jun 8, 2026

Uncontrolled Hemorrhagic Shock Modeled via Liver Laceration in Mice with Real Time Hemodynamic Monitoring
Published on: May 21, 2017
Efficacy of different hemostatic devices for severe liver bleeding: a randomized controlled animal study
Ildikó Takács1, Jürgen Wegmann, Szabolcs Horváth
1University of Pécs, Pécs, Hungary.
Background:
Correct hemostasis in liver surgery is hard to achieve because of the oozing bleeding. The aim of this study was to compare the potential benefits of a new compress to the 2 commercial hemostatic compresses.
Methods:
Collagen- and cellulose-based hemostatics were investigated. A standardized resection was treated by applying different hemostatics in a randomized order, and bleeding times were measured. Macroscopic evaluation of the liver and tissue sampling for histological investigations were carried out after 21 days.
Results:
The bleeding times of bovine collagen (BoCo), protein-coated equine collagen (PECo), and oxidized cellulose (OxCe) were 140 ± 88, 243 ± 140 (P = .005 vs BoCo), and 352 ± 70 s (P < .001 vs BoCo), respectively. Microscopic evaluation of the PECo presented fibrosis and significant inflammation in the implantation zone, whereas BoCo and OxCe caused only fibrosis in the wound area.
Conclusion:
BoCo showed significantly better hemostatic effect than PECo and OxCe.