Related Experiment Video
Updated: Jul 23, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Comparison of a TLC method with EMIT and GC/MS for detection of cannabinoids in urine
1Center for Human Toxicology, University of Utah, Salt Lake City 84112.
Abstract:
Urine specimens were analyzed in parallel with a new TLC method, an EMIT assay, and a reference GC/MS method. At a 9-carboxy-THC cutoff of 20 ng/mL, the TLC method correctly identified 92% of the positive urines and 97% of the negative urines. In contrast, only 63% of the urine specimens shown by GC/MS to contain greater than 20 ng/mL of 9-carboxy-THC were identified as positive by the EMIT d.a.u. assay at the 100-ng/mL cannabinoid cutoff.