[A realistic cardiovascular model for evaluating clinical applicability of duplex and color-Doppler equipment]
F Brassel1, P Hendrickx, U Roth
1Medizinische Hochschule Hannover, Abteilung Diagnostische Radiologie II im Krankenhaus Oststadt.
Abstract:
To evaluate the accuracy and reliability of colour-coded Doppler devices (CCDD) for practical use, a circulatory phantom with in vivo characteristics was developed. The experimental design consists of a piston-diaphragm metering pump with an adjustable piston volume and frequency of stroke as well as silicone tubings with different elasticity and diameters comparable with cadaver arteries. Depending on physiological conditions, blood as a circulating fluid and the surrounding soft tissues of the artery near the Doppler probe should be employed. The transducer should be manually guided. Thus, comparable to physiological frequency spectral displays, different types of wave forms (e.g. monophasic, triphasic) were initiated to analyze the reliability of CCDDs for practical use.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...