Related Experiment Video
Updated: Jun 8, 2026

06:46
A High-Throughput Multiplexed Screening for Type 1 Diabetes, Celiac Diseases, and COVID-19
Published on: July 5, 2022
A one-step immunoassay for carbohydrate antigen 19-9 by biosensor based on imaging ellipsometry
Chenghong Huang1, Yanyan Chen, Gang Jin
1Suzhou Institute of Nano-tech and Nano-bionics, Chinese Academy of Sciences, Suzhou, 215123, China. huang1ng@tom.com
Annals of Biomedical Engineering
|October 1, 2010
Summary
A novel biosensor using imaging ellipsometry (BIE) enables a rapid, one-step immunoassay for carbohydrate antigen 19-9 (CA19-9) cancer marker. This method shows high accuracy and reproducibility, indicating potential for clinical diagnostics.
Area of Science:
- Biomedical Engineering
- Analytical Chemistry
- Immunotechnology
Background:
- Accurate cancer marker detection is crucial for early diagnosis and management.
- Existing methods for cancer biomarkers like carbohydrate antigen 19-9 (CA19-9) can be time-consuming.
- Development of simple, rapid, and reliable diagnostic assays is essential for clinical settings.
Purpose of the Study:
- To develop and evaluate a one-step immunoassay for CA19-9 using biosensor-based imaging ellipsometry (BIE).
- To assess the analytical performance, including sensitivity, specificity, reproducibility, and accuracy, of the developed BIE immunoassay.
- To investigate the potential clinical applicability of the BIE immunoassay for CA19-9 detection.
Main Methods:
- Immobilization of anti-CA19-9 monoclonal antibody on a carboxyl-modified silicon substrate.
- Blocking of the sensing layer with bovine serum albumin (BSA).
- Detection of CA19-9 using biosensor-based imaging ellipsometry (BIE).
Main Results:
- The BIE immunoassay demonstrated a linear detection range of 15.0–120.0 U/mL with a limit of detection of 10.0 U/mL (S/N=3).
- High reproducibility was observed with intra-slide and inter-slide coefficients of variation (CV%) below 10%.
- Accuracy ranged from 95.1-106.7%, specificity was high (cross-reaction rate <5.8%), and results correlated strongly with ECLIA (R²=0.997) in clinical samples.
Conclusions:
- The developed one-step BIE immunoassay provides a sensitive, specific, and reproducible method for CA19-9 detection.
- The assay exhibits excellent accuracy and strong correlation with established methods, suggesting clinical utility.
- Biosensor-based imaging ellipsometry offers a promising platform for rapid and efficient cancer biomarker analysis in clinical diagnostics.