Related Experiment Video
Updated: Jul 28, 2026

Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
Output characteristics of DeVilbiss No. 40 hand-held jet nebulizers
K N Chan1, M M Clay, M Silverman
1Dept of Paediatrics and Neonatal Medicine, Royal Postgraduate Medical School Hammersmith Hospital, London, UK.
Abstract:
DeVilbiss No. 40 hand-held nebulizers are widely used for quantifying airway responsiveness in large populations using pharmacological agents. We examined the aerosol characteristics of five nebulizers. Within each device, the aerosol output and droplet size were reasonably stable over a wide range of bulb pressures, although there were considerable differences in output characteristics between nebulizers. The droplet size was very large compared to conventional aerosol delivery systems, with a mass median diameter greater than 10 microns for three of the five devices. Between 28-50% of the output was in particles sufficiently small for airway deposition (less than 6.2 microns). A more vigorous compression of the bulb caused a small increase output and a reduction in droplet size, resulting in a much bigger variation in the output of the respirable aerosol (less than 6.2 microns) with changes in bulb pressure. The loss due to evaporation was about 3.5%, causing a similar rise in the osmolality of the nebulizer solution. In view of the variable nebulizer output and the marked between-operator variation in bulb pressure, the characteristics of individual DeVilbiss No. 40 nebulizers should be evaluated by individual operators before use in clinical practice or research.
More Related Videos
Related Concept Videos
Inhaled Medications
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Suctioning the Nasopharyngeal Airway
Equipment Required
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Tracheostomy Suctioning II: Procedure