Evaluation of a new pediatric continuous oximetry catheter
Hadi Mohseni-Bod1, Helena Frndova, Rose Gaitaro
1Departments of Critical Care Medicine, The Hospital for Sick Children, Toronto, Canada. hadi.mohseni-bod@sickkids.ca
Purpose:
To evaluate the accuracy of central venous oxygen saturation recordings from a new in-line pediatric oximetry catheter.
Design:
Prospective, observational study.
Study Population:
Eighteen pediatric patients who needed central venous access for monitoring and/or treatment between January 2006 and June 2006 in the pediatric intensive care unit of the Hospital for Sick Children in Toronto, Canada.
Methods And Main Results:
Measurements were done at the baseline and then every 4-8 hrs. The monitor was calibrated in vivo at the baseline and then daily. In vitro calibration of the monitor was also performed in the last five patients. The hemoglobin value was updated when there was a significant change. The maximum duration of sampling was 72 hrs (if indicated). There were 131 measurements in 17 patients; each subject had a different number of paired measurements (median 5). Three patients were excluded due to violation of the protocol, and 113 measurements were left in analysis. The mean difference of catheter value from the laboratory value was -1.01 (median 0). The interquartile range was 5. The difference of both methods was evenly distributed as per a Bland-Altman plot, with one patient's data lying outside of the comparable limits of ± 1.96 sd from the mean differences. The relationship of the difference between the catheter data and the lab data to the independent variables (age, weight, gender, catheter tip, diagnosis, and signal quality index) was estimated by using the multiple regression analysis (version 9.1, SAS Institute, Cary, NC). All variables were eliminated. The Pearson correlation coefficient between lab-mixed venous oxygen saturation and oximetry catheter readings for measurements was 0.88.
Conclusion:
In this limited number of patients, use of the PediaSat venous oximetry catheter was safe and had good agreement with co-oximetry-measured values.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.