[Digital volume tomography (DVT) and multislice spiral CT (MSCT): an objective examination of dose and image quality]
Y Kyriakou1, D Kolditz, O Langner
1Institut für Medizinische Physik, Friedrich-Alexander Universität Erlangen-Nürnberg, Erlangen.
Purpose:
In the last five years digital volume tomographs (DVT) have found their way into the diagnostic imaging of the facial skull. In this study both the image quality and dose of DVT and multislice spiral CT (MSCT) in this field of application were investigated using established physical methods for CT.
Materials And Methods:
Measurements on DVT scanners of various manufacturers and on a modern MSCT scanner were performed. The investigation was based on equivalent dose levels for both modalities (CT dose index, CTDI). For this purpose, the dose was measured with an ionization chamber in a cylindrical PMMA phantom. For the evaluation of image quality, the spatial resolution, contrast and noise were investigated with phantoms established for CT.
Results:
MSCT exhibited spatial resolution values of 1.0 to 1.6 lp/mm, while DVT provided resolution between 0.6 and 1.0 lp/mm only. Thus, MSCT offered similar or better resolution at an equivalent dose. For soft tissue resolution, DVT showed significant image artifacts. MSCT yielded higher homogeneity and no significant artifacts, and the contrast steps of the phantom were more verifiable. The different DVT devices, from image intensifiers to modern flat-detector (FD) devices, showed significant differences in favor of the FD devices.
Conclusion:
For medium and high contrast applications (teeth/bones), DVT scanners can be an alternative to MSCT at comparable radiation exposure. However, MSCT offers advantages in terms of constantly good and controlled image quality with significantly more flexible scan parameters at a constant or lower dose and should therefore be given preference.
More Related Videos
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...