Related Experiment Video
Updated: Jun 8, 2026

Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
An in vitro investigation of image-guided steerable catheter navigation
1State Key Laboratory of Robotics and System, Harbin Institute of Technology, Room 417, Bio-X Centre, Bldg 2E, Science Park, 2 Yikuang Street, Nangang District, Harbin, Heilongjiang 150080, People's Republic of China. liuhao.hit@gmail.com
Abstract:
The objective of this study is to provide catheterization with a three-dimensional (3D) guiding image and reduce the collision probability between the catheter tip and vascular wall. A bidirectional steerable catheter was integrated with two magnetic position-tracking sensors on both sides of its bending segment. The tracking information was displayed on the guiding image, which helped the surgeons to determine the relative position between the catheter tip and surrounding vessels. The navigation path was generated on the basis of the vascular skeleton. Moreover, along the path, a series of guiding circular planes were set as the guidance for the catheter. Three operations (bending, advancing, and twisting) were jointly conducted to get the catheter through these guiding planes in turn and eventually into the target vessel. The effectiveness of the proposed navigation method was verified by experiments implemented in an aorta vascular phantom. The navigation system has a mean error of 0.19 mm, a root mean square of 0.49 mm, and a standard deviation of 0.46 mm.

