Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
Thrombogenic potential of Innovia polymer valves versus Carpentier-Edwards Perimount Magna aortic bioprosthetic
Thomas E Claiborne1, Gaurav Girdhar, Siobhain Gallocher-Lowe
1Department of Biomedical Engineering, Stony Brook University, Stony Brook, New York 11794-8181, USA.
Abstract:
Trileaflet polymeric prosthetic aortic valves (AVs) produce hemodynamic characteristics akin to the natural AV and may be most suitable for applications such as transcatheter implantation and mechanical circulatory support (MCS) devices. Their success has not yet been realized due to problems of calcification, durability, and thrombosis. We address the latter by comparing the platelet activation rates (PARs) of an improved polymer valve design (Innovia LLC) made from poly(styrene-block-isobutylene-block-styrene) (SIBS) with the commercially available Carpentier-Edwards Perimount Magna Aortic Bioprosthetic Valve. We used our modified prothrombinase platelet activity state (PAS) assay and flow cytometry methods to measure platelet activation of a pair of 19 mm valves mounted inside a pulsatile Berlin left ventricular assist device (LVAD). The PAR of the polymer valve measured with the PAS assay was fivefold lower than that of the tissue valve (p = 0.005) and fourfold lower with flow cytometry measurements (p = 0.007). In vitro hydrodynamic tests showed clinically similar performance of the Innovia and Magna valves. These results demonstrate a significant improvement in thrombogenic performance of the polymer valve compared with our previous study of the former valve design and encourage further development of SIBS for use in heart valve prostheses.