Related Experiment Video
Updated: Jun 8, 2026

Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Utilization of Electrocardiographic P-wave Duration for AV Interval Optimization in Dual-Chamber Pacemakers
Dan Sorajja1, Mayurkumar D Bhakta, Luis Rp Scott
1Division of Cardiovascular Diseases, Mayo Clinic Arizona, 5777 E Mayo Blvd, Phoenix, AZ 85054, USA.
Background:
Empiric programming of the atrio-ventricular (AV) delay is commonly performed during pacemaker implantation. Transmitral flow assessment by Doppler echocardiography can be used to find the optimal AV delay by Ritter's method, but this cannot easily be performed during pacemaker implantation. We sought to determine a non-invasive surrogate for this assessment. Since electrocardiographic P-wave duration estimates atrial activation time, we hypothesized this measurement may provide a more appropriate basis for programming AV intervals.
Methods:
A total of 19 patients were examined at the time of dual chamber pacemaker implantation, 13 (68%) being male with a mean age of 77. Each patient had the optimal AV interval determined by Ritter's method. The P-wave duration was measured independently on electrocardiograms using MUSE® Cardiology Information System (version 7.1.1). The relationship between P-wave duration and the optimal AV interval was analyzed.
Results:
The P-wave duration and optimal AV delay were related by a correlation coefficient of 0.815 and a correction factor of 1.26. The mean BMI was 27. The presence of hypertension, atrial fibrillation, and valvular heart disease was 13 (68%), 3 (16%), and 2 (11%) respectively. Mean echocardiographic parameters included an ejection fraction of 58%, left atrial index of 32 ml/m(2), and diastolic dysfunction grade 1 (out of 4).
Conclusions:
In patients with dual chamber pacemakers in AV sequentially paced mode and normal EF, electrocardiographic P-wave duration correlates to the optimal AV delay by Ritter's method by a factor of 1.26.
Related Concept Videos
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...