Related Experiment Video
Updated: Jun 8, 2026

Improvement of a Closed Chest Porcine Myocardial Infarction Model by Standardization of Tissue and Blood Sampling Procedures
Published on: March 12, 2018
A nonimaging catheter for measurement of coronary artery lumen area: a first in man pilot study
James Hermiller1, Jenny S Choy, Mark Svendsen
1The Heart Center of IN, Indianapolis, Indiana, USA.
Objectives:
The objective of this human pilot study was to determine the safety and the level of agreement between a novel nonimaging 2.7 Fr. catheter-based system (LumenRECON, LR) that uses electrical conductance for measurement of lumen cross-sectional area (CSA) with intravascular ultrasound (IVUS) and quantitative coronary angiography (QCA). Based on previous animal studies, we hypothesized the level of agreement between LR and IVUS to be 13%.
Background:
Accurate and reproducible vessel sizing is essential for optimal percutaneous coronary intervention (PCI).
Methods:
A total of 12 patients were studied to evaluate the safety, accuracy, and reproducibility of the system in comparison with IVUS and QCA. The CSA of coronary arteries was determined by IVUS, QCA, and LR in the distal, proximal, and center of a lesion during standard PCI.
Results:
A Bland-Altman plot of the LR versus IVUS and QCA show a nonsignificant mean difference between the two measurements of 0.04 and 0.07 mm in diameter, respectively. The root mean square error of LR versus IVUS and QCA was 14.3 and 25.8% of the mean IVUS or QCA diameter, respectively. The mean of the difference between two LR duplicate measurements was nearly zero (0.03 mm) and the repeatability coefficient was within 8.7% of the mean of the two measurements. There were no procedural complications nor were any device-related MACE reported within 30 days of the procedure.
Conclusions:
This proof of concept pilot study establishes the safety and accuracy of the conductance technology for a pivotal trial of coronary sizing.
More Related Videos
08:51Intra-cardiac Side-Firing Light Catheter for Monitoring Cellular Metabolism using Transmural Absorbance Spectroscopy of Perfused Mammalian Hearts
Published on: May 12, 2019
06:39Novel Percutaneous Approach for Deployment of 3D Printed Coronary Stenosis Implants in Swine Models of Ischemic Heart Disease
Published on: February 18, 2020
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Acute Coronary Syndrome III: Diagnostic Studies