Related Experiment Video
Updated: Jun 8, 2026

Orthopedic Robot-Assisted Femoral Neck System in the Treatment of Femoral Neck Fracture
Published on: March 3, 2023
New orthopedic devices and the FDA
Ujash Sheth1, Nhu-An Nguyen, Sean Gaines
1Queen's School of Medicine, Queen's University, Hamilton, Ontario, Canada. guyklein1@yaho.com
Abstract:
Each year the field of orthopedics is introduced to an influx of new medical devices. Each of these medical devices has faced certain hurdles prior to being approved for marketing by the U.S. Food and Drug Administration (FDA). Among the regulatory pathways available, the 510(k) premarket notification is by far the one most commonly used. The 510(k) premarket notification allows the manufacturer to receive prompt approval of their device by demonstrating that it is "substantially equivalent" to an existing legally marketed device. In most instances, this proof of substantial equivalence allows manufacturers of medical devices to bypass the use of clinical trials, which are a hallmark of the approval process for new drugs. As a result, most medical devices are approved without demonstrating safety or effectiveness. This article reviews the regulatory processes used by the FDA to evaluate new orthopedic devices.
Related Concept Videos
FDA Approved Drugs: Changes to Approved Drugs
Drug Products: Biologics, Biosimilars and Interchangeables
Drug Regulation