Biosensor for Detection of Antibiotic Resistant Staphylococcus Bacteria
Published on: May 8, 2013
Evaluation of molecular assays for rapid detection of methicillin-resistant Staphylococcus aureus
Surbhi Malhotra-Kumar1, Liesbet Van Heirstraeten, Andie Lee
1Department of Medical Microbiology, Vaccine & Infectious Disease Institute, Campus Drie Eiken, University of Antwerp, S3, Universiteitsplein 1, B-2610 Wilrijk, Antwerp, Belgium. surbhi.malhotra@ua.ac.be
Abstract:
The diagnostic sensitivities of the BD GeneOhm and Cepheid Xpert assays were compared using culture on log-serial dilutions of well-characterized methicillin-resistant Staphylococcus aureus (MRSA) and non-MRSA strains and on nasal and groin swabs from patients with histories of MRSA carriage. The sensitivities of GeneOhm and Xpert were high at 10(3)-CFU/ml MRSA concentrations (92.3% and 96.3%, respectively) although decreased considerably (<35%) at a 1-log-lower concentration. Unexpectedly, both assays also detected select coagulase-negative staphylococci, which requires further evaluation.