AngioSculpt PTCA Balloon Catheter entrapment and detachment managed with stent jailing

Reji Pappy1, Archana Gautam, Mazen S Abu-Fadel

  • 1University of Oklahoma Health Sciences Center, and Veteran Affairs Medical Center, Cardiovascular Section, Oklahoma City, Oklahoma, USA.

Insights

The first reported case of AngioSculpt EX catheter entrapment during percutaneous coronary intervention (PCI) was successfully managed. Stent jailing secured the device, preventing further complications in the right coronary artery.

Area of Science:

  • Cardiovascular Interventions
  • Medical Device Technology

Background:

  • The AngioSculpt PTCA Scoring Balloon Catheter is an FDA-approved device utilized for plaque modulation during percutaneous coronary intervention (PCI).
  • Catheter entrapment is a rare but significant complication during PCI procedures.

Observation:

  • This report details the first documented instance of AngioSculpt EX catheter entrapment and detachment within the right coronary artery.
  • The event occurred during a PCI procedure aimed at plaque modulation.

Findings:

  • The entrapped AngioSculpt EX catheter and its shaft were successfully managed using a technique called stent jailing.
  • This method involved securing the device against the vessel wall, resolving the entrapment complication.

Implications:

  • This case highlights a novel management strategy for AngioSculpt EX catheter entrapment during PCI.
  • The successful stent jailing technique provides a potential solution for similar rare device-related complications in complex coronary interventions.