Microcontroller-based system for estimate of calcium in serum samples
Abstract:
In this study, a microcontroller-based control unit was designed and constructed for the estimation of serum calcium in blood samples. The proposed optoelectronic instrument used a red light emitting diode (LED) as a light source and photodiode as a sensor. The performance of the system was compared with that of a commercial instrument in measuring calcium ion. The quantitative analysis of calcium in a catalyst using arsenazo III as colorimetric reagent was used to test the device. The calibration curve for calcium binding with arsenazo III was drawn to check the range of linearity, which was between 0.1 to 4.5 mM L⁻¹. The limit of detection (LOD) is 0.05 mM L⁻¹. Absorbance changes over the pH range of 2-12 were determined to optimize the assay, with maximum absorption at pH 9.0. Interferences in absorbance from monovalent (K+ and Na+) and divalent (Mg²+) cations were also studied. The results show that the system works successfully.