A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Arrhythmia discrimination by physician and defibrillator: importance of atrial channel
Igor Diemberger1, Cristian Martignani, Mauro Biffi
1Institute of Cardiology, University of Bologna, Azienda Ospedaliera S.Orsola-Malpighi, Bologna, Italy. igor.diemberger@gmail.com
Background:
Many ICD carriers experience inappropriate shocks, but the relative merits of dual- /single-chamber devices for arrhythmia discrimination still remain unclear. We explored possible advantages of the atrial data provided by dual-chamber implantable defibrillators (ICD) for discrimination of real-life supraventricular/ventricular tachyarrhythmias (SVT/VT).
Methods:
100 dual-chamber traces from 24 ICD were blindly reviewed in dual-chamber and simulated single-chamber (with/without discriminator data) reading modes by five electrophysiologists who determined chamber of origin and provided Likert-scale "confidence" ratings. We assessed 1) intra/interobserver concordance; 2) diagnostic accuracy, using expert diagnoses as a reference standard; 3) ROC curves of sensitivity/specificity of "likelihood perception" scores, generated by combining chamber-of-origin diagnostic judgments with Likert-scale "confidence" ratings. We also assessed diagnostic accuracy of automated discrimination by all possible dual-/single-chamber algorithm configurations.
Results:
Interobserver concordance was "substantial" (modified Cohen kappa-test values for dual-/single-chamber, 0.79/0.68); intraobserver concordance "almost complete" (kappa ≥ 0.89). Dual-chamber mode provided best diagnostic sensitivity/specificity (99%/92%) and highest reader confidence (p<0.001). Area under ROC curves of sensitivity/specificity values for the "likelihood perception" score (representing electrophysiologists' perceptions of the likelihood that an episode was of ventricular origin) was highest in dual-chamber mode (0.98 vs. 0.93 for both single-chamber modes; p<0.001). Regarding automated discrimination, all four dual-chamber configurations conferred 100% sensitivity (specificity values ranged 39%-88%), whereas single-chamber configurations appeared inferior (best sensitivity/specificity combination, 89%/64%).
Conclusions:
Availability of the atrial channel helps in reducing inappropriate ICD therapies by providing relevant advantages in terms of both appropriate cardiologist's post-hoc discrimination of SVT/VT (improving program tailoring) and automated arrhythmia discrimination.
Related Concept Videos
Dysrhythmias V: Evaluating Dysrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation III: AED Use
Dysrhythmias II: Classification of Tachyarrhythmias