Related Experiment Video
Updated: Jun 8, 2026

Validated LC-MS/MS Panel for Quantifying 11 Drug-Resistant TB Medications in Small Hair Samples
Published on: May 19, 2020
Rational design for variability minimization in bioanalytical method validation: illustration with LC-MS/MS assay
Sanjay Gurule1, Arshad Khuroo, Tausif Monif
1Department of Clinical Pharmacology and Pharmacokinetics, Ranbaxy Laboratories Ltd, Plot no. GP-5, HSIDC, Sector-18, Udyog Vihar Industrial Area, Old Delhi Gurgaon Road, Gurgaon 122015, Haryana, India.
Abstract:
Terbinafine, a widely used antifungal drug, is a challenging molecule for quantitative bioanalysis due to certain factors contributing assay variability. Despite previous attempts at human plasma determination of terbinafine, exhaustive stability of the drug or an internal standard was lacking. Internal standard stability with negligible variation throughout the analysis is an indicator of a reliable bioanalytical method as the majority of LC-MS/MS assays are based on analyte/IS response ratios for quantitation. A newly developed high-throughput simple LC-MS/MS method is described for human plasma determination of terbinafine using naftifine internal standard and eluting all compounds within 2 min. A solid-phase extraction of terbinafine achieving mean recovery of 84.3% (CV < 4%) without compromising sensitivity (limit of quantitation 5.11 ng/mL) or linearity (5.11-3014.19 ng/mL) is delineated in this paper. A heated nebulizer in positive multiple reaction monitoring mode was employed with transitions m/z 292.2 →141.1 and 288.2 →117.0 for terbinafine and naftifine, respectively, resulting in excellent chromatographic separation on a Hypurity Advance (50 x 4.6 mm, 5 µm) column. The developed method was successfully applied to clinical samples and for the first time demonstrated marked improved extraction efficiency and reliable long-term plasma stability results without any internal standard response variation during the entire course of study.
Related Concept Videos
Data Validation
Key parameters for method validation include:
Measurement of Bioavailability: Pharmacodynamic Methods
Measurement of Bioavailability: Pharmacokinetic Methods