Identification of Rare Bacterial Pathogens by 16S rRNA Gene Sequencing and MALDI-TOF MS
Published on: July 11, 2016
Rapid identification of staphylococci from prosthetic joint infections using MALDI-TOF mass-spectrometry
Llinos G Harris1, Khalid El-Bouri, Stuart Johnston
1Medical Microbiology and Infectious Diseases, Institute of Life Science, School of Medicine, Swansea University, Swansea, United Kingdom. l.g.harris@swansea.ac.uk
Abstract:
Hospital-acquired infections associated with implanted medical devices are most commonly caused by staphylococci. Current methods of species identification are slow, costly, and sometimes unreliable. We evaluated the ability of a Bruker Daltonics Microflex MALDI-TOF/MS in conjunction with MALDI Biotyper software to identify 158 characterized staphylococcal isolates from prosthetic joint infections, including 36 Staphylococcus aureus, 100 Staphylococcus epidermidis, 10 Staphylococcus capitis, 8 Staphylococcus lugdunensis, 2 Staphylococcus warneri, and 2 Staphylococcus haemolyticus isolates using the extraction method recommended by Bruker Daltonics. The suggested species identification by the MALDI Biotyper software was correct for all isolates, indicating reliable differentiation between S. aureus and coagulase-negative staphylococci. Applying the recommended criteria of the MALDI Biotyper software all 158 isolates gave scores ≥2.0, implying secure genus and probable species identification for all isolates. 34/36 S. aureus, 36/100 S. epidermidis, 5/10 S. capitis, 6/8 S. lugdunensis, 2/2 S. haemolyticus, 0/2 S. warneri displayed scores ≥2.3 implying highly probable species identification. For S. epidermidis 25/100 additional isolates had a score close to 2.3. It appears that additional clinically relevant staphylococcal isolates in the data base might aid in identification at scores implying highly probable species identification. The ability of the MALDI Biotyper software to recognize clonally-related strains within a species group (i.e. sub-typing) was investigated, and showed great potential. In conclusion, the MALDI-TOF/MS MALDI Biotyper system provides a promising rapid and reliable method of identifying clinical isolates from prosthetic joint infections to the species level, and has potential for sub-typing.
Related Concept Videos
Rapid Identification of Pathogens
MALDI-TOF Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Matrix-Assisted Laser Desorption Ionization (MALDI)
Automated Microbial Diagnostics