Purpose:
In moving target irradiation with pencil beam scanning, the interplay effect between the target motion and the scanned beam is a problem because this effect causes over or under dosage in the target volume. To overcome this, we have studied rescanning using a gating technique.
Methods:
A simulation and experimental study was carried out. In the experiment, we used the fast scanning system developed at the HIMAC to verify the validity of phase controlled rescanning method, in which the time for rescanning irradiation of each slice is matched to the gating duration.
Results:
Simulation and experimental results showed that controlling the scan speed to match the respiration cycle with rescans can deliver the blurred dose distribution. In the comparison between the static measurements and the moving measurements with the phase controlled rescanning method, the dose difference was less than 2% for pinpoint chambers in the target volume.
Conclusions:
The simulation and experimental results demonstrated that the phase controlled rescanning method makes it possible to deliver the dose distribution close to the expected one. As an experimental result for 3D irradiation, it was estimated that blurring by the probability density function was not only for a lateral distribution, but also for a distal distribution, even in the lateral rescanning.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...