Related Experiment Video
Updated: Jun 7, 2026

Murine Intrapulmonary Tracheal Transplantation: A Model for Investigating Obliterative Airway Disease After Lung Transplantation
Published on: November 10, 2023
Management of recurrent airway strictures in lung transplant recipients using AERO covered stents
Justin H Tan1, Nicholas Fidelman, Jeremy C Durack
1Department of Radiology, University of California, San Francisco, CA 94143, USA.
Abstract:
The potential role of AERO tracheobronchial covered stents in the management of recurrent postinfectious strictures of the bronchus intermedius was studied in three lung transplant recipients. Six devices were inserted. Five of the stents migrated immediately on placement. Buildup of thick mucus was observed in all stents remaining in the airway for longer than 1 week. Strictures recurred in all patients 1, 3, and 5 months after stent deployment. Attempts at stent retrieval were successful for three of five devices. The use of AERO stents may not offer a therapeutic advantage versus balloon dilation of bronchus intermedius strictures in lung transplant recipients.