Four-dimensional lung treatment planning in layer-stacking carbon ion beam treatment: comparison of layer-stacking and conventional ungated/gated irradiation
Related Experiment Video
Updated: Jun 7, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Shinichiro Mori1, Nobuyuki Kanematsu, Hiroshi Asakura
1Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan. shinshin@nirs.go.jp
Purpose:
We compared four-dimensional (4D) layer-stacking and conventional carbon ion beam distribution in the treatment of lung cancer between ungated and gated respiratory strategies using 4DCT data sets.
Methods And Materials:
Twenty lung patients underwent 4DCT imaging under free-breathing conditions. Using planning target volumes (PTVs) at respective respiratory phases, two types of compensating bolus were designed, a full single respiratory cycle for the ungated strategy and an approximately 30% duty cycle for the exhalation-gated strategy. Beams were delivered to the PTVs for the ungated and gated strategies, PTV(ungated) and PTV(gated), respectively, which were calculated by combining the respective PTV(Tn)s by layer-stacking and conventional irradiation. Carbon ion beam dose distribution was calculated as a function of respiratory phase by applying a compensating bolus to 4DCT. Accumulated dose distributions were calculated by applying deformable registration.
Results:
With the ungated strategy, accumulated dose distributions were satisfactorily provided to the PTV, with D95 values for layer-stacking and conventional irradiation of 94.0% and 96.2%, respectively. V20 for the lung and Dmax for the spinal cord were lower with layer-stacking than with conventional irradiation, whereas Dmax for the skin (14.1 GyE) was significantly lower (21.9 GyE). In addition, dose conformation to the GTV/PTV with layer-stacking irradiation was better with the gated than with the ungated strategy.
Conclusions:
Gated layer-stacking irradiation allows the delivery of a carbon ion beam to a moving target without significant degradation of dose conformity or the development of hot spots.