Estimation of the minimum dose required to measure ventricular width in follow-up cranial computed tomography (CCT)

K Kirchhof1, W A Wohlgemuth, A Berlis

  • 1Institut und Poliklinik für Radiologische Diagnostik, Universitätsklinikum Carl Gustav Carus, Dresden, Germany. Klaus.Kirchhof@uniklinikumdresden.de

Rofo : Fortschritte Auf Dem Gebiete Der Rontgenstrahlen Und Der Nuklearmedizin
|October 26, 2010
PubMed

Insights

Radiation dose for pediatric hydrocephalus imaging can be reduced significantly using cranial computed tomography (CCT) without compromising ventricular width measurements. Standard protocols with reduced tube current (100 kV) are generally safe.

Area of Science:

  • Pediatric Radiology
  • Medical Imaging Physics

Background:

  • Hydrocephalus in children often requires serial cranial computed tomography (CCT) scans to monitor ventricular width.
  • Optimizing radiation dose during CCT is crucial to minimize cumulative exposure in pediatric patients.

Purpose of the Study:

  • To determine the minimum radiation dose for CCT that reliably measures ventricular width in pediatric hydrocephalus.
  • To assess the impact of varying tube current and voltage on measurement accuracy.

Main Methods:

  • A phantom simulating infant calvarium and lateral ventricles was scanned using two multi-slice CT scanners.
  • Scans were performed with varied tube current (100-400 mA) and voltage (80-140 kV).
  • Ventricular width was measured and compared to reference values obtained at higher doses.

Main Results:

  • Significant radiation dose reduction (up to 85%) was achieved with both scanners without affecting measurement reliability.
  • Specific combinations of tube voltage and current led to less reliable measurements despite higher doses.

Conclusions:

  • No single dose or setting guarantees optimal reliability and minimal exposure for all scanners.
  • Reducing tube current to 100 kV within standard protocols is a safe guideline for reliable ventricular width assessment.
Abstract