Objective:
To evaluate the review criteria for complete blood analysis.
Methods:
total of 8820 blood samples taken at our hospital were detected both by the Sysmex XE-2100 automated blood cell analyzer and double-blind method microscopy. The performance of two review criteria were evaluated by statistical analysis. The parameters, such as true and false positive rates, true and false negative rates, sensitivity, specificity, positive predictive value, negative predictive value, total effective rate, test effectiveness and re-examination rate were calculated according to the criteria of XE-2100 hematology analyzer re-examination standards coordination group of Society of Laboratory Medicine of Chinese Medical Association. And a study was conducted to determine the improved clinical effectiveness over the criteria.
Results:
The values of true and false positive rates, true and false negative rates, sensitivity, specificity, positive predictive values, negative predictive value, total effective rate, test effectiveness and reexamination rate were 13.0%, 27.2%, 55.3%, 4.5%, 74.3%, 67.0%, 32.4%, 92.5%, 68.3%, 21.2%, and 40.2% respectively. The re-examination rate decreased to 33.8% and the total effective rate increased to 74.7% according to our improved review criteria. And there was no missed diagnosis of leukemic cells. It was found that 2069 samples (23.5%) needed manual microscopy for WBC differential analysis, 847 had abnormal cells and no leukemia was missed.
Conclusions:
The automatic blood cell analyzer and WBC differential analysis review criteria have great clinical utilities. The review criteria of blood cell analysis may be improved according to patient source and instrument performance so as to better meet the needs of clinical application.
Related Concept Videos
Blood Transfusion
Blood Transfusion Overview
A blood transfusion is a medical procedure used to replace blood lost due to injury, surgery, or to treat conditions such as anemia or cancer. During a transfusion, donor blood is...
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...