Related Experiment Video
Updated: Jun 7, 2026

Diffusion Imaging in the Rat Cervical Spinal Cord
Published on: April 7, 2015
[Influence of imaging parameters on the measurement of apparent diffusion coefficient]
Masanori Ozaki1, Akio Ogura, Isao Muro
1School of Allied Health Sciences, Kitasato University.
Abstract:
The apparent diffusion coefficient (ADC) values are calculated by using signal intensity in diffusion-weighted images (DWIs) with two or more different b-value. Therefore, the signal to noise ratio (SNR) of DWI with higher b-value may have a big influence on the measured ADC value. We examined the influence of the imaging parameters on the calculated ADC values. The SNR of DWI increased by using a larger voxel size, by means of a decreased number of matrix, an increased slice thickness, and an increased field of view (FOV). However, when the number of excitations was increased to improve the SNR of DWI, the signal intensity of background noise was observed to be slightly increased. It was suggested that the consistency of measured ADC was not preserved when the signal of the DWIs with higher b-value dropped close to the noise level.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Imaging Studies VII: Vascular Imaging
Imaging Studies III: Computed Tomography

