Related Experiment Video
Updated: Jun 7, 2026

Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
Reliability of a digital image method for measuring medial mandibular flexure in dentate subjects
Rosemary Sadami Arai Shinkai1, Simone de Andrade Canabarro, Caroline Bom Schmidt
1Department of Prosthodontics, Pontifical Catholic University of Rio Grande do Sul, Porto Alegre, Brazil.
Unlabelled:
Medial mandibular flexure (MMF) is the functional approximation of hemi-mandibles in jaw opening and protrusion, which may affect conventional or implant-supported arch prostheses.
Objective:
This study evaluated the intra- and inter-rater reliability of a digital image method for linear measurement of MMF in dentate subjects.
Material And Methods:
Mandibular occlusal impressions in vinyl polysiloxane during rest and maximum opening were obtained from seven adult volunteers. Each impression and the Mitutoyo digital calipers with a 10mm opening were scanned at a 200% magnification. The images were processed using Adobe Photoshop software to get reference points on contralateral first molars. Linear intermolar measurements were taken using the Image Tool software (UTHSCSA), which was calibrated with the standard opening of the digital calipers. Intermolar distance was measured in triplicate for each image by two independent examiners. MMF was calculated by subtracting the intermolar distance in maximum opening (or protrusion) from the distance in rest position. Data were analyzed by intraclass correlation (ICC) tests for estimating intra- and inter-rater reliability.
Results:
MMF values ranged from -0.21 to 0.44mm. Intra-rater ICCs were 0.982 and 0.993, and inter-rater ICC was 0.696.
Conclusion:
The present results show excellent intra-rater and good inter-rater reliability of this digital image method for measuring MMF in dentate subjects.