Related Experiment Video
Updated: Jun 7, 2026

A Spine Robotic-Assisted Navigation System for Pedicle Screw Placement
Published on: May 11, 2020
[Pedicle screw fixation of thoracic spinal fracture assisted by CT-based navigation system]
Dong Zhou1, Nan-Wei Xu, Lu-Ming Nong
1Department of Orthopedics, Changzhou No. 2 Affiliated Hospital of Nanjing Medical University, Changzhou 210003, China. zhoudong1012@hotmail.com
Objective:
To evaluate the feasibility and accuracy of thoracic pedicle screw fixation when assisted by a CT-based navigation system.
Methods:
A total of 102 thoracic pedicle screws inserted by a CT-based navigation system (n = 24) (Group A) and 98 thoracic pedicle screws inserted by X-ray fluoroscopy (n = 22) (Group B) were observed by postoperative CT scan. The accuracy of both series was compared. And the process of navigation was investigated.
Results:
Among Group A, 99 were of grade I (97.1%), 3 grade II and 0 grade III; among Group B, 88 were of grade I (89.8%), 8 grade II and 2 grade III. One patient of grade III had the intractable pain of chest and back. Neural damage was not demonstrated. The difference was statistically significant (P < 0.05).
Conclusion:
A CT-based navigation system can increase accuracy of thoracic pedicle screw fixation.