Related Experiment Video
Updated: Jun 7, 2026

Aptamer-Based Target Detection Facilitated by a 3-Stage G-Quadruplex Isothermal Exponential Amplification Reaction
Published on: October 6, 2022
Ultrasensitive electrochemical aptasensor for thrombin based on the amplification of aptamer-AuNPs-HRP conjugates
Jie Zhao1, Youyu Zhang, Haitao Li
1Key Laboratory of Chemical Biology and Traditional Chinese Medicine Research (Ministry of Education), College of Chemistry and Chemical Engineering, Hunan Normal University, Changsha 410081, PR China.
Abstract:
Successful development of an ultrasensitive and highly specific electrochemical aptasensor for thrombin based on amplification of aptamer-gold nanoparticles-horseradish peroxidase (aptamer-AuNPs-HRP) conjugates was reported. In this electrochemical protocol, aptamer1 (Apt1) was immobilized on core/shell Fe(3)O(4)/Au magnetic nanoparticles (AuMNPs) and served as capture probe. Aptamer2 (Apt2) was dual labeled with AuNPs and HRP and used as detection probe. In the presence of thrombin, the sandwich format of AuMNPs-Apt1/thrombin/Apt2-AuNPs-HRP was fabricated. Remarkable signal amplification was realized by taking the advantage of AuNPs and catalytic reactions of HRP. Other proteins, such as human serum albumin, lysozyme, fibrinogen, and IgG did not show significant interference with the assay for thrombin. Linear response to thrombin concentration in the range of 0.1-60 pM and lower detection limit down to 30 fM (S/N=3) was obtained with the proposed method. This electrochemical aptasensor is simple, rapid (the whole detection period for a thrombin sample is less than 35 min), sensitive and highly specific, it shows promising potential in protein detection and disease diagnosis.