Related Experiment Video
Updated: Jun 7, 2026

An Experimental Protocol for Femtosecond NIR/UV - XUV Pump-Probe Experiments with Free-Electron Lasers
Published on: October 23, 2018
Two-chord interferometry using 3.39 μm He-Ne laser on a flux-coil-generated FRC
1Tri Alpha Energy, Inc., Rancho Santa Margarita, California 92688, USA. hgota@trialphaenergy.com
Abstract:
A two-chord λ(IR)∼3.39 μm He-Ne laser interferometer system was developed for a flux-coil-generated field-reversed configuration to estimate the electron density and the total temperature of the field-reversed configuration (FRC) plasma. This two-chord heterodyne interferometer system consists of a single ∼2 mW infrared He-Ne laser, a visible (λ(vis)∼632.8 nm) He-Ne laser for the alignment, a 40 MHz acousto-optic modulator, photodetectors, and quadrature phase detectors. Initial measurement was performed and the measured average electron densities were 2-10×10(19) m(-3) at two different radial positions in the midplane. A time shift in density was observed as the FRC expands radially. The time evolution of the line-averaged density agrees with the density estimated from the in situ internal magnetic probes, based on a rigid-rotor profile model.